當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 215S二氯苯C6H4Cl2氣體檢測管 215S p-DICHLOROBENZENE 10-150ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
二氯苯C6H4Cl2氣體檢測管 215S p-DICHLOROBENZENE 10-150ppm
1. PERFORMANCE
1) Measuring range :10-150 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :6 ppm
4) Shelf life :1 year (Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Blueish purple
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By reacting with an Oxidizer, Hydrogen chloride is prodaced and PH indicator is discoloured.
C6H4CI2+PbO2+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
150
120
100
80
60
40
20
10
ppm
200ml
????
Tube No.
215S p-DICHLOROBENZENE
95
Substance Interference Coexistence
Benzene Similar stains are produced and higher
readings are given.
Toluene 〃
Blueish green stains are produced and
Hexane the top of discoloured layer becomes 〃
unclear.
Ethanol The accuracy of reading is not affected. The accuracy of reading is not affected.
TEMPERATURE CORRECTION TABLE
Tube Corrected Concentration (ppm)
Readings 10℃ 15℃ 20℃ 25℃ 30℃ 35℃ 40℃
(ppm) (50°F) (59°F) (68°F) (77°F) (86°F) (95°F) (104°F)
150 - - 150 108 66 48 30
100 - - 100 72 46 37 26
50 110 80 50 28 26 21 16
20 50 27 20 14 10 8 6
10 22 15 10 8 6 5 4