當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 232SA乙二醇HOCH2CH2OH氣體檢測管232SA ETHYLENE GLYCOL20-250mg/m3
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
乙二醇HOCH2CH2OH氣體檢測管232SA ETHYLENE GLYCOL20-250mg/m3
Tube No.232SA
1. PERFORMANCE
1) Measuring range :20-250mg/m3
Number of pump strokes :2(200ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :5mg/m3
4) Shelf life :2 years
5) Operating temperature :20~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 2 pump strokes
8) Colour change :Pink→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Formic acid is produced and PH indicator is discoloured.
HOCH2CH2OH+HIO4→2HCHO+HIO3+H2O
HCHO+HIO4+H2SO4→HCOOH+HIO3
HCOOH+NaOH→Na(HCOO)+H2O
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
DETECTOR TUBE
(White)
PRETREAT TUBE
250
200ml
mg/m3
50
20
200
150
100
232SA ETHYLENE GLYCOL
Substance Interference Coexistence
Aldehydes Similar stain is produced. Higher readings are given.
Sulphur dioxide 〃〃
Ethylene oxide 〃〃
Hydrogen sulphide Orange/ Pink stain is produced. Accurate measurement can not
be done.
TEMPERATURE CORRECTION TABLE
Scale True Concentration (mg/m3)
Readings 20℃ 22℃ 25℃ 30℃ 40℃
(mg/m3) (68°F) (71.6°F) (77°F) (86°F) (104°F)
250 - 370 250 200 155
200 - 260 200 165 130
150 270 170 150 125 105
100 120 110 100 85 75
50 60 55 50 45 40
30 23 20 20 18 15
請輸入賬號
請輸入密碼
請輸驗證碼
以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,化工儀器網對此不承擔任何保證責任。
溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。