當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 139SB乙酸正丙酯CH3CO2(CH2)2CH3氣體檢測管 139SB PROPYL ACETATE
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
乙酸正丙酯CH3CO2(CH2)2CH3氣體檢測管 139SB PROPYL ACETATE 0.01-1.4%
Tube 39SB
1. PERFORMANCE
1) Measuring range :0.01-1.4%
Number of pump strokes 2(200ml)
2) Sampling time :4 minutes/2 pump strokes
3) Detectable limit :10 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Graduations printed on the tube are calibrated by Methyl ethyl ketone at 2 pump
strokes and Isobutyl acetate concentration is determined by using a conversion chart.
7) Colour change :Orange→Pale brownish green
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:5%
3. CHEMICAL REACTION
Potassium dichromate is reduced
CH3CO2CH2CH(CH3)2+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
1.4
1.2
1.0
0.8
0.6
0.4
0.3
0.2
0.1
0.05
0.01
%
139SB C ISOBUTYL ACETATE
39SB Tube reading (%)?
Isobutyl acetate(%)?
1.0
0.8
0.6
1.4
1.2
0.4
0.2
0.1
0.05
0.01 0.01
0.1
0.2
0.5
1.0
1.4
(200mR)?
Substance Interference ppm Coexistence
Acetylene 3% Whole reagent is changed to Pale
brown.
Propane 0.2% 〃
Other organic gases or vapours Similar stain is produced.
except Halogenated hydrocar- 50 Higher readings are given.