當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 163SA環氧丙烷氣體檢測管163SA PROPYLENE OXIDE 0.05-5.0%
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
環氧丙烷氣體檢測管163SA PROPYLENE OXIDE 0.05-5.0%
Tube 63SA
1. PERFORMANCE
1) Measuring range :1.0-5.0% 0.05-3.0%
Number of pump strokes 1/2(50ml) 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :20 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Orange→Dark brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Potassium dichromate is reduced.
CH3CHCH2O+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
3.0
2.5
2.0
1.5
1.0
0.5
0.2
0.05
0.1
%
GAS To pump
DEHUMIDIFIER
(White)
163SA PROPYLENE OXIDE
0
2
3
1
Toluene〔%〕?
FIG.1 Influence of Toluene
Propylene oxide indicating value 〔%〕
0.25 0.5
Interference
0.05
0.1
0.2
0.5
1.0
1.5
2.0
2.5
3.0
5.0
4.0
3.0
2.0
1.0
Conversion scale
(50mR actual value)?
(100mR concentration scale)?
(%)?
0
2
3
1
Ethyl acetate〔%〕?
FIG.2 Influence of Ethyl acetate
Propylene oxide indicating value 〔%〕
1 2 0
2
3
1
Methanol〔%〕?
FIG.3 Influence of Methanol
Propylene oxide indicating value 〔%〕
0.5 1.0
Substance Interference ppm Coexistence
Aromatic hydrocarbons FIG.1 Similar stain is produced. Higher readings are given.
Esters FIG.2 〃〃
Ketones 〃〃
Alcohols FIG.3 〃〃
Halogenated hydrocarbons WBrhoowlen .reagent is changed to 0.5% 〃
(NOTE)
In case of 1 / 2 pump strokes, following conversion scale is available for actual concentration.