當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 158S苯乙烯C6H5CH=CH2氣體檢測管158S STYRENE 2.5-300ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
苯乙烯C6H5CH=CH2氣體檢測管158S STYRENE 2.5-300ppm
Tube 58S
1. PERFORMANCE
1) Measuring range :5-300 ppm 2.5-150 ppm
Number of pump strokes 1(100ml) 2(200ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :0.5 ppm(200ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high : 10%
3. CHEMICAL REACTION
A polymer of Styrene is produced by sulphuric acid.
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
300
250
200
150
100
50
20
5
ppm
GAS To pump
158S STYRENE
Substance ppm Interference ppm Coexistence
Acrylonitrile FIG.1 Tafhfeec atecdc.uracy of readings is not 400 Lower readings are given.
Butadiene 3 Similar stain is produced. 5 Uneven discolouration is produced
FIG.2 and higher readings are given.
Formaldehyde 〃15 Yellowish orange stain is produced
FIG.3 and higher readings are given.
Acetaldehyde 〃350 Similar stain is produced and
higher readings are given.
Methyl alcohol The accuracy of readings is not 0.35% Pale discolouration is produced
affected. and higher readings are given.
Ethyl alcohol 〃0.18% 〃
Ethyl acetate 〃700 〃
Butyl acetate 〃700 〃
(NOTE)
In case of 2 pump strokes, following formula is available for actual concentration.
Actual concentration=1/2×Reading value.
0
200
100
Acrylonitrile Butadiene
FIG.1 Influence of Acrylonitrile Styrene indicating value 〔
ppm〕
500 1000 0
200
100
FIG.2 Influence of Butadiene
Styrene indicating value 〔ppm〕
〔ppm〕? 50 100 0
200
100
Formaldehyde
FIG.3 Influence of Formaldehyde