1. PERFORMANCE
1) Measuring range :10-300 ppm 5-150 ppm
Number of pump strokes 1/2(50ml) 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :1 ppm(100ml)
4) Shelf life :2 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Red
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Hydrogen chloride is produced and PH indicator is discoloured.
CCI2=CCI2+PbO2+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
150
100
100
30
20
10
5
ppm
GAS To pump
PRETREAT REAGENT
(Grey)
Tube No.
135SA TETRACHLOROETHYLENE
Substance Interference Coexistence
Vinyl chloride Similar stain is produced. Higher readings are given.
Hydrogen chloride FIG.1 〃〃
1,2-Dichloroethylene FIG.2 〃〃
Trichloroethylene 〃〃
Chlorine Pale red stain is produced.
(NOTE)
In case of 1 /2 or 2 pump strokes, following formula is available for the actual concentration.
Actual concentration=2×Temperature corrected value
0
200
100
Hydrogen chloride〔ppm〕?
FIG.1 Influence of Hydrogen chloride
Tetrachloroethylene
indicating value 〔ppm〕
500 1000
Coexistence
Interference
0
200
100
1,2-Dichloroethylene〔ppm〕?
FIG.2 Influence of 1,2-Dichloroethylene
Tetrachloroethylene
indicating value 〔ppm〕
50 100
Coexistence
Interference
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
150 - 172 150 138 134
100 144 116 100 92 88
50 70 56 50 46 44
30 40 36 30 28 26
20 22 21 20 19 18
10 10 10 10 10 10